Skip to content

Burn-in software setup

The software for the burn-in is divided into to parts:

  1. The sensor readout and hardware control are running on the Raspberry Pi
  2. The data collection of sensor readouts and data visualization are on Yipeng's desktop computer (referred as Julian in the following sections)

Installation

After clone the NeoBurnIn to the pi, run nix-shell in the project root. This will automatically install all required dependencies.

Note

The pi must be running on NixOS for nix-shell to work!

Note

After nix-shell finishes executing, you'll be dropped in a shell. Only in this shell one can run burn-in related scripts.

Scripts to run on the Raspberry Pi

To start the burn-in testing one should issue the command tmux and then open three separate panels using the key combination Ctrl+A+". One can open more panels if needed although three panels should be sufficient to run the burn-in:

  1. In the first panel, run CtrlServer.py using the command ./CtrlServer.py --config-file ./config/burn_in_ctrl_server.yml

  2. In the second panel, run CtrlClient.py using the command ./CtrlClient.py --config-file ./config/burn_in_ctrl_client.yml

  3. The third panel is reserved for interactive use, such as controlling the MARATON or USB relay.

Note

To navigate between tmux panels, type Ctrl+A+up/down arrow keys in order to move up/down between panels.

To stop the burn-in testing, navigate to each tmux panel which is running and type Ctrl+C in order to issue a keyboard interrupt which will stop the code from running. After all codes are stopped, one can exit tmux panels by issuing the command exit within each panel.

Warning

The CtrlClient.py should be stoped after a burn-in is finished, to avoid collecting idle MARATON current readouts.

Scripts to run on Julian

Yipeng will take care of that. The gist is that DataServer.py and another data visualization script will be running on Julian

Controlling the PSU (MARATON) with curl

CtrlServer.py is required to control PSU remotely.

To turn on/off a MARATON channel, use the commands

curl -X POST http://192.168.1.30:45679/psu/192.168.1.31/<CHANNEL_NUMBER>/on
curl -X POST http://192.168.1.30:45679/psu/192.168.1.31/<CHANNEL_NUMBER>/off

Note

<CHANNEL_NUMBER> is the number of the MARATON channel which one would like to turn on/off.

There are 12 MARATON channels that can be turned on/off, ranging from channel 1 to channel 12.

If changing the MARATON channel on/off is successful then the terminal will output Success and the LEDs on the LVRs for that MARATON channel will turn on/off respectively.

Warning

WEINER control software on Windows numbers MARATON channel in a different way: Starting from 0 to 11 (instead of 1 to 12).

This means that issuing the curl command to turn on channel 1 will turn on the channel labeled 0 on the WEINER control software. If one is not using the WEINER control software to monitor the MARATON channels then do not worry about this numbering offset.

Controlling the USB relay with curl

CtrlServer.py is required to control PSU remotely.

To turn on/off a USB relay channel, use the commands after checking the dynamic address

curl -X POST http://192.168.1.30:45679/relay/0001:0004:00/<CHANNEL_NUMBER>/on
curl -X POST http://192.168.1.30:45679/relay/0001:0004:00/<CHANNEL_NUMBER>/off

Note

<CHANNEL_NUMBER> is the number of the USB relay channel which one would like to turn on/off.

There are only two USB relay channels, channel 1 and channel 2, and only one channel will need to be turned on/off.

This is because only one channel will be connected to a solenoid valve for the burn-in (as of 3/26/20). To see which channel one must turn on/off with curl commands one must inspect the hardware setup of the system, although the hardware should be setup such that the solenoid valve is connected to channel 2 of the USB relay.

Warning

The number 0001:0014:00 in the curl commands can change and one should check if this number has changed before issuing these curl commands. To check whether this number has changed:

curl -X POST http://127.0.0.1:45679/relay/list